Description
The HI96719 is a portable magnesium hardness photometer that combines accuracy and ease of use in an ergonomic and portable design. The user can accurately determine the magnesium hardness concentration within a range of 0.00 to 2.00 mg/L (ppm) using the prepared HI93719-0 reagents. The HI96719 offers many advanced features including the unique CAL Check ™ function that is used for efficiency verification and meter calibration.
Absorbance Specifications:
– Light source: Tungsten lamp
– Light detector: Silicon photocell with narrowband interference filter @ 525 nm
Additional Specifications:
– Range: 0.00 to 2.00 mg/L (ppm)
– Resolution: 0.01 mg/L
– Accuracy @ 25°C (77°F): ±0.11 mg/L ±5% of reading.
– Method: Adaptation of Standard Methods for Drinking Water and Wastewater Analysis, 18th Edition, EDTA Colorimetric Method
– Auto Power Off: After ten minutes of inactivity in measurement mode; after one hour of inactivity in calibration mode; with reminder of the last reading.
– Environmental conditions: 0 to 50 °C (32 to 122 °F); RH max 95% non-condensing
– Weight: 360g (12.7oz.)
– Battery Type: 9V Battery
– Dimensions: 193 x 104 x 69 mm (7.6 x 4.1 x 2.7″)