Description
Portable Fluoride Meter – HI98402
The HI98402 is a portable, waterproof and rugged fluoride meter. This meter offers many features including automatic calibration, on-screen instructions, and automatic temperature compensation. The HI98402 requires a half-cell fluoride electrode with HI4010 epoxy body and a HI5313 reference electrode constructed of PEI material. They are available as accessories.
The HI98402 is a rugged, portable fluoride meter designed to provide laboratory accuracy and accuracy in harsh environmental conditions. This meter is ideal for a variety of applications in drinking water treatment, environmental monitoring, and electroplating baths. The HI98402 offers a variety of features including five distinct measurement ranges, automatic calibration to 1 or 2 points, automatic temperature compensation, and on-screen instructions.
The HI98402 uses a BNC input to connect the fluoride half-cell electrode and another for the reference electrode. A separate RCA connector is used for a thermistor temperature probe. The temperature probe provides measurement with accuracy of ±0.2°C for accurate temperature compensation.
General Features
One- or Two-Point Calibration – Automatic calibration can be performed at 1 or 2 points using the 5 programmed standards (1 mg/L, 2 mg/L, 10 mg/L, 100 mg/L, and 1000 mg/L).
On-Screen Instructions – Clear and easy-to-use on-screen instructions will guide you through all the steps to perform a calibration and take measurements.
Fluoride Half Cell Electrode and Reference – The HI98402 requires two half-cell electrodes to measure fluoride. The HI4010 is a fluoride half-cell electrode that includes an epoxy body and lanthanum fluoride crystal sensing membrane.
The HI5313 is a half-cell reference electrode with a tapered junction that produces a highly stable reference potential and constant electrolyte flow. The gauge, electrodes and temperature probe assembly form a complete and accurate measurement system.
Automatic Temperature Compensation – The HI98402 offers automatic temperature compensation for fluoride measurements in the range of -5.0 to 55.0°C.
Specs:
Temperature Range: -20.0 to 120.0°C (-4.0 to 248.0°F)
Temperature Resolution: 0.1°C (0.1°F)
Temperature Accuracy: ±0.2°C; ±0.4°F (excluding probe error)
Temperature Compensation: Automatic, -5 to 55 °C (with temperature probe).
Fluoride range: 0.050 to 0.500 mg/L (ppm); 0.50 to 5.00 mg/L (ppt) 5.0 to 50.0 mg/L; 50 to 500 mg/L; 0.50 to 1.90 g/L (ppt).
Fluoride Resolution: 0.001 mg/L (ppm); 0.01 mg/L; 0.1 mg/L; 1 mg/L; 0.01 g/L
Fluoride Accuracy: ±5% of reading or ±0.02 mg/L (ppm) fluoride (at ± 3°C calibration temperature).
Fluoride Calibration: Automatic at one or two points at 1 mg/L, 2 mg/L, 10 mg/L, 100 mg/L and 1000 mg/L
Required Probes: HI4010 fluoride electrode with BNC connector and 1 m (3.3′) cable (not included); HI5313 reference electrode with 1 m (3.3 ‘) cable (not included), HI7662 stainless steel temperature probe with 1 m (3.3’) cable (not included).
Battery Type/Duration: 1.5V AAA (3) / approximately 200 hours of continuous use.
Ambient: 0 to 50°C (32 to 122°F); Max HR 100%
Dimensions: 185 x 72 x 36 mm (7.3 x 2.8 x 1.4 “”)
Weight: 300 g (10.6 oz.)